AquaBplus Water Purification System, Part Number 24-2000-0, AquaBplus 2500 - Typ USA (G02040108-US), AquaBplus 2000 - Typ USA (G02040114-US), AquaBplus HF -Typ USA (G02040120FMC-US)
FDA Recall Z-1364-2022 · Posted None
Advertisement
Recalling FirmFresenius Medical Care Holdings, Inc.
Recall StatusTerminated
Product CodeFIP
Date Initiated2021-08-03
Date Terminated2024-04-10
Product Quantity2 units
Reason for Recall
AquaBplus HF module on the AquaBplus 2000 osmosis system, an error in the US Installation Guideline instructed that 3-ring mains, instead of 2-ring mains, may be installed in combination with an AquaBplus 2000 device. The 3-ring main and fluid fly loops the heat disinfection may not be as effective as a 2-ring main installation
Firm Action
Customers were notified via telephone beginning 8/3/21. An FMCRTG, LLC Water Systems Specialist contacted the two consignees via phone and scheduled a time to perform the inspection on the AquaBplus 2000 devices. No formal written communication will be sent since calls and inspections have already performed.
Root Cause
Labeling Change Control
Distribution Pattern
US Nationwide distribution in the states of NC and WV.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.