Monaco Radiation Treatment Planning System, Version 3.10.00. Planning of radiation therapy.

FDA Recall Z-1365-2012 · Posted 2012-04-03

Recall StatusTerminated
Product CodeMUJ
Date Initiated2012-03-05
Date Terminated2014-04-01
Product Quantity44

Reason for Recall

An unnecessary dose refresh can occur in certain situations which causes inconsistencies in the Beam Doses and the Total Dose.

Firm Action

An "IMPORTANT SAFETY NOTICE" was issued between 3/5-9/12 via e-mail, FAX, or traditional mail. The Notice described the correct functioning of the system, the problem, when and why the issue occurs, the clinical impact, and the workaround. Patch 3.10.01 is now available. Customers can download their patch at the link provided in the Notice. The Notice also provided a list of offices where customers can contact if they have questions or further support is needed.

Root Cause

Software design

Distribution Pattern

Worldwide Distribution -- USA, including the states of OH, IN, WI, CA, FL, IL, LA, NJ, MA, NC, and MO and the countries of Canada, Australia, Belgium, China, France, Germany, Japan, The Netherlands, New Zealand, Portugal, Switzerland, Taiwan, and the United Kingdom.