Terminated

Arrow PICC powered by Arrow VPS Precision Stylet Product Code:CDC-45052-VPS2 Product Usage: The Arrow Pressure Injectable PICC is indicated for short-term or longterm peripheral access to the central venous system for intravenous therapy, blood sampling, infusion, pressure injection of contrast media and allows for central venous pressure monitoring. The maximum pressure of pressure injector equipment used with the pressure injectable PICC may not exceed 300 psi.

FDA Recall Z-1365-2019 · Posted None

Recall StatusTerminated
Product CodeLJS
Date Initiated2018-12-13
Date Terminated2020-10-01
Product Quantity963 units (US) in total

Reason for Recall

Kits labeled as containing a 40 cm PICC may contain a 50 cm PICC, and kits labeled as containing a 50 cm PICC may contain a 40 cm PICC

Firm Action

Arrow International initiated recall notification letter on 13th December 2018 via FedEx 2-day mail describing problem ,health risk and action to take:, immediately discontinue use and quarantine the products. Complete the enclosed Acknowledgement Form as soon as possible indicating whether you do or do not have stock and fax it to 1-855-419-8507 or email to [email protected]. A customer service representative will contact you with a Return Goods Authorization (RGA) number and provide instructions for the return of products.

Root Cause

Employee error

Distribution Pattern

US Nationwide Distribution