Draeger Globe-Trotter and Globe-Trotter IC, Model numbers MS20508 and MS20509. neonatal transport system
FDA Recall Z-1365-2022 · Posted None
Advertisement
Recalling FirmDraeger Medical Systems, Inc.
Recall StatusOpen, Classified
Product CodeFPL
Date Initiated2022-05-20
Product Quantity38 OUS (no US distribution)
Reason for Recall
The firm has initiated this recall to replace component on devices distributed containing a component within the oxygen supply system that is not rated to the working pressure of the system.
Firm Action
On May 20, 2022, the firm began to send Important Field Safety Notice letters to affected customers. Customers were instructed to avoid using the ventilator portion of the system due to the potential for enriched ambient air rising beyond safe levels until a replacement part can be installed and tested.
The firm will send customers a replacement kit and testing which will be completed at no charge to the customer. Draeger Service Representatives will call customers and schedule a service visit to update affected systems.
Root Cause
Under Investigation by firm
Distribution Pattern
No US distribution. Foreign distribution to BY, CL, DE, ES, GB, HU, IN, IT, KW, MX, SA, SV, TW.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.