AccuData GTS Plus for Blood Glucose Monitoring and Automated Data Management including GTS Plus with Accu-Chek Advantage blood glucose meter, Cat no. 3000249 and/or GTS with Accu-Chek Advantage blood glucose meter, Cat. no. 404, and/or Replacement GTS with Accu-Chek Advantage blood gulcose meter, Cat. no. 404R, and/or Replacement Base Unit with Advantage Module, Cat. no. 3426050. Ref. nos. 030002490001 and 03136833001 both have meters with model no. 777.

FDA Recall Z-1366-06 · Posted 2006-08-05

Recall StatusTerminated
Product CodeCGA
Date Initiated2006-06-22
Date Terminated2008-04-07
Product Quantity55,189 meters (for all Recalls Z-1354-1371-06)

Reason for Recall

The meter gives an error message that can "actually" mean either a problem with the strip or a blood glucose too low to measure, but the meter error message only reports that there is a bad strip.

Firm Action

A recall notification letter dated April 24, 2006 was sent to each customer alerting them to the fact that this error code has a dual meaning. New software will be installed in all meters when it becomes available.

Root Cause

Other

Distribution Pattern

Worldwide, USA, Germany, Guam, North Mariana Islands, Panama and Virgin Islands.