Proxima Drape labeled as MINOR PACK and LAPAROSCOPY. Sterile surgical drape.
FDA Recall Z-1366-2025 · Posted 2025-03-14
Advertisement
Recalling FirmMEDLINE INDUSTRIES, LP - Northfield
Recall StatusOpen, Classified
Product CodeFDE
Date Initiated2025-02-13
Product Quantity506 units
Reason for Recall
Potential breach in pouch packaging which could lead to loss of sterility.
Firm Action
Medline distributed an IMMEDIATE ACTION REQUIRED MEDICAL DEVICE RECALL notification, dated 2/13/25, to consignees who purchased individually packaged Proxima sterile surgical drapes and kits that contained Proxima sterile surgical drapes.
Consignees that purchased kits containing recalled product are to immediately check stock for affected kits and place them in quarantine. Consignees are to complete the provided response form to receive over-labels to be applied to quarantined kits with additional instructions to remove the affected components prior to use of the kit.
Consignees that purchased individual units of Proxima sterile surgical drapes are to immediately check stock for affected devices and immediately destroy identified product. Consignees are to complete the provided response form to obtain credit for the destroyed product.
If a consignee is a distributor or have further distributed affected devices, they are instructed to forward the provided recall notification. Questions about this recall can be directed to Medline via phone at 866-359-1704 or by email to [email protected].
Root Cause
Packaging process control
Distribution Pattern
US Nationwide distribution.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.