Terminated

Gambro cartridge blood tubing set used with Gambro Dasco Phoenix and Cobe Centrysystem 3 (and 3+) Hemodialysis System, Gambro, Lakewood, CO 80401

FDA Recall Z-1367-2008 · Posted 2008-03-25

Recall StatusTerminated
Product CodeKDI
Date Initiated2007-09-10
Date Terminated2012-01-03
Product Quantity41,000,000

Reason for Recall

Kinked tubing on hemodialysis device may cause hemolysis.

Firm Action

On September 10, 2007, the firm sent an Advisory Notice to all consignees alerting them that a kink in the tubing can result in physical hemolysis. The firm included with the Notice a Quick Reference Guide that provides drawing of the proper installation of the Phoenix and describes how to prevent bloodline kinking. A customer reply form was also included. The firm is in the process of making labeling changes that will be released with the updated software revisions.

Root Cause

Device Design

Distribution Pattern

Nationwide, including 17 VA facilities and 1 military facility. Foreign distribution to Canada, Korea, Mexico.