Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

YC-1800 OPHTHALMIC YAG LASER SYSTEM; The YC-1800 allows the safe and bloodless treatment of intraocular diseases for outpatients by mechanical disruption caused by energy of the YAG laser pulses.

FDA Recall Z-1367-2015 · Posted 2015-04-02

Recalling FirmNidek Inc
Recall StatusTerminated
Product CodeLXS
Date Initiated2015-03-06
Date Terminated2015-09-03
Product Quantity799 units

Reason for Recall

Dear Doctor letter and Quick Reference Guide is being sent to users of the YC-1800 to reiterate important safety and cleaning information after complaints of pitting lens that may cause unclear images.

Firm Action

Dear Doctor letters were sent along with a Quick Reference guide to all affected sites in February.

Root Cause

Process control

Distribution Pattern

Nationwide Distribution.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑