EvoTech Endoscope Cleaner & Reprocessor, Product Code 50004 Indicated for use with high level disinfectant CIDEX OPA Concentrate and an enzymatic detergent, CIDEZYME GI, to achieve cleaning and high level disinfection of heat sensitive semi-critical endoscopes.

FDA Recall Z-1368-2008 · Posted 2008-08-08

Recall StatusTerminated
Product CodeFEB
Date Initiated2008-02-13
Date Terminated2011-11-29
Product Quantity64 units

Reason for Recall

Cleaning cycles are being cancelled. Specific issues are: a) Water heater tank is rusting resulting in the in-line 0.2 micron filter prematurely clogging, b) Interface board malfunction resulting in a premature lid opening, and c) Minor fluid leaks within the system (commonly at the 0.2 micron filter attachment points).

Firm Action

Feburary 13, 2008 ASP issued an Urgent Product Recall Letter. ASP recommends that customers use an alternative reprocessing method until they return the system. However, if they do not have an alternate method for processing endoscopes and choose to continue using the EvoTech ECR System during this time, endoscopes processed through completed cycles in accordance with the User's Guide will be appropriately cleaned and highlevel disinfected. ASP is prepared to tailor an alternative solution to minimize the interruption to the customer's processing needs.

Root Cause

Other

Distribution Pattern

Worldwide Distribution --- including USA and countries of Canada and Japan.