Power Processor 3060 Tube Refrigerated Stockyard, P/N 6915556.

FDA Recall Z-1369-05 · Posted 2005-08-13

Recalling FirmBeckman Coulter Inc
Recall StatusTerminated
Product CodeLXG
Date Initiated2005-05-05
Date Terminated2012-03-07
Product Quantity46

Reason for Recall

Device does not maintain a uniform internal temperature within 2-8 C at warmer ambient temperatures.

Firm Action

Recall letters mailed 5/5/2005 request users monitor the temperature variances and evaluate for continued use in thier laboratory.

Root Cause

Process change control