EvoTech Endoscope Cleaner and Reprocessor, Product Code 50004-002 Indicated for use with high level disinfectant CIDEX OPA Concentrate and an enzymatic detergent, CIDEZYME GI, to achieve cleaning and high level disinfection of heat sensitive semi-critical endoscopes.
FDA Recall Z-1369-2008 · Posted 2008-08-08
Advertisement
Recalling FirmAdvanced Sterilization Products
Recall StatusTerminated
Product CodeFEB
Date Initiated2008-02-13
Date Terminated2011-11-29
Product Quantity2 units
Reason for Recall
Cleaning cycles are being cancelled. Specific issues are:
a) Water heater tank is rusting resulting in the in-line 0.2 micron filter prematurely clogging,
b) Interface board malfunction resulting in a premature lid opening, and
c) Minor fluid leaks within the system (commonly at the 0.2 micron filter attachment points).
Firm Action
Feburary 13, 2008 ASP issued an Urgent Product Recall Letter. ASP recommends that customers use an alternative reprocessing method until they return the system. However, if they do not have an alternate method for processing endoscopes and choose to continue using the EvoTech ECR System during this time, endoscopes processed through completed cycles in accordance with the User's Guide will be appropriately cleaned and highlevel disinfected. ASP is prepared to tailor an alternative solution to minimize the interruption to the customer's processing needs.
Root Cause
Other
Distribution Pattern
Worldwide Distribution --- including USA and countries of Canada and Japan.