Allura XPER FD 20 and and Allura CV20; Allura XPER FD 10 and Field Extension for XPER Cardiovascular System and Allura CV20. Angiographic X-ray systems with a solid state X-ray imaging device for cardiovascular and vascular diagnostic and interventional procedures.

FDA Recall Z-1369-2011 · Posted 2011-03-22

Recall StatusTerminated
Product CodeIZI
Date Initiated2010-11-04
Date Terminated2011-07-21
Product Quantity7 units in US

Reason for Recall

Adjustment of the micro-switches in the footswitch used on the Philips cardiovascular X-ray systems was not implemented. If it is programmed for exposure, it is possible that after pressing the pedal on the footswitch it does not release itself. The patient may be exposed to X-ray longer than intended.

Firm Action

On 11/5/10, Philips began sending the Electronic Product Radiation Warning notification letter, dated November 4, 2010, to their consignees. The consignees were informed that the middle pedal of the foot switch of the Allura FD10, FD20, CV20 systems, and the Allura Field Extension series does not not always become inactive (sticks) when pedal is released. The function of the middle pedal on the foot switch is freely programmable by the customer. If it is programmed for exposure and if the pedal does not release by itself, the patient will be exposed to X-ray longer than intended. The consignees were advised to check if their footswitch is possibly affected. Footswitch identification 452270000141 is noted. If the pedal does not release by itself press the red emergency stop button on the system. Consignees can call Philips Healthcare Call Center at 800-722-9377, #5 and Reference FCO 72200189.

Root Cause

Pending

Distribution Pattern

Worldwide Distribution -- US, including states of CA, IL, MA, NC, and OC.