Terminated

ACCU-CHEK Connect Diabetes Management App; Instruction Manual Designed to transfer data for diabetes management.

FDA Recall Z-1369-2015 · Posted 2015-04-02

Recall StatusTerminated
Product CodeJQP
Date Initiated2014-10-30
Date Terminated2015-06-08
Product QuantityThe application has been downloaded total of 644 times (Italy-219, South Africa-24, Germany-401) and there are 113 bolus advice activations.

Reason for Recall

Roche Diabetes Care has become aware of an issue with the Accu-Chek Connect diabetes management app that could potentially lead to inaccurate bolus advice being provided to the user. A thorough investigation of the situation revealed that this issue may occur if the user changes the screen orientation of the phone from portrait to landscape or vise versa while looking at the Bolus Advisor or Carb

Firm Action

Urgent Field Safety Notices were sent to the customer and healthcare professional starting on 10/30/2014. The letter provided a description of the reason for the notification; and recommended that all users should update their app immediately.

Root Cause

Software design

Distribution Pattern

There are no US distributors as the affected device is not marketed in the US. The affected device was distributed in Germany, Italy and South Africa. The application has been downloaded total of 644 times (Italy-219, South Africa-24, Germany-401) and there are 113 bolus advice activations.