Helix Elite" Inactivated Standard: Inactivated SARS-CoV-2 Whole Virus (Pellet), Multi-analyte, unassayed quality control material, Catalog Number HE0065N
FDA Recall Z-1371-2025 · Posted 2025-03-14
Advertisement
Recalling FirmMicrobiologics Inc
Recall StatusOpen, Classified
Product CodeOHQ
Date Initiated2025-02-20
Product Quantity6 units
Reason for Recall
The A549 human cell target (human cells for sample adequacy control) may give a late Ct value.
Firm Action
An Urgent Medical Device Recall notification letter dated 2/20/25 was sent to customers.
INSTRUCTIONS:
1. REVIEW lab procedures to understand how this information affects your usage
2. DISCARD affected product.
3. COMPLETE the response form provided.
4. RETURN the response form to [email protected]
5. KEEP this letter for your records.
6. NOTIFY other facilities or customers to whom the product may have been distributed or transferred.
7. CONTACT Microbiologics if a replacement kit is needed.
Microbiologics is dedicated to ensuring the success of your quality control processes. Please contact our Recall Support team at 320.229.7080 or [email protected] with questions or concerns. Collect calls may be made.
Root Cause
Process control
Distribution Pattern
US Nationwide distribution in the states of TX. UK.
Official Source