Hex Driver Part Number HXDRIVER-2.0 AO; Hex Driver Part Number HXDRIVR-2.5 AO; Hex Driver Part Number HXDRIVR-2.5CL AO; Hex Driver Part Number DVHX-2.5/130C AO; Hex Driver Part Number HXDRIVR-3.0 AO; Hex Driver Part Number HXDRIVER-3.5AO
FDA Recall Z-1373-2018 · Posted None
Advertisement
Recalling FirmTriMed Inc.
Recall StatusTerminated
Product CodeLXH
Date Initiated2018-01-11
Date Terminated2024-01-18
Product Quantity1525 devices
Reason for Recall
Following the completion of sterilization cycle(s) of the impacted Hex Driver, the epoxy color band on the impacted Hex Driver can migrate down the shaft and/or potentially onto the sterilization tray or caddy.
Firm Action
Recall notification letters were sent to affected consignees on January 16, 2018. The consignee letter requests that the consignee return an initial Acknowledgement &
Receipt Form to TriMed by January 23, 2018.
Root Cause
Process control
Distribution Pattern
US, China, Latin America