ProSpecT Giardia/Cryptosporidium Microplate Assay. Catalog #: 2458496.
FDA Recall Z-1375-04 · Posted 2004-08-17
Advertisement
Recalling FirmRemel, Inc.
Recall StatusTerminated
Product CodeMHI
Date Initiated2003-12-09
Date Terminated2004-09-15
Product Quantity127 kits
Reason for Recall
The conjugate may cause the negative control to produce high OD readings, therefore invalidating the run.
Firm Action
Remel contacted each customer with a letter, by mail, requesting a count of the affected lots to be returned. Product would be replaced upon customer request.
Root Cause
Other
Distribution Pattern
US and OUS customers.