Biomet Vanguard DCM CR Tibial Bearing, 18MM x 71/75MM, CR STD bearing, sterile, Biomet Orthopedics, Inc., Warsaw, IN; Part 183448.

FDA Recall Z-1375-2008 · Posted 2008-07-25

Recalling FirmBiomet, Inc.
Recall StatusTerminated
Product CodeJWH
Date Initiated2008-02-12
Date Terminated2009-12-22
Product Quantity4

Reason for Recall

Mislabeled: The packaging of the two products was mixed up, resulting in the products being labeled with an incorrect size. If implanted, the product may not fit properly, which could require repeat surgery. The problem could also result in a delay in the procedure.

Firm Action

Distributors were notified via recall letter (Urgent Medical Device Removal Notice) dated 2/12/08 to locate the devices in their possession and at hospital sites, discontinue use, and return them to Biomet. Hospitals having received the devices will be provided with a recall letter dated 2/12/08 which describes the nature of the problem and the removal of these instruments. A third letter dated 2/12/08 was provided for implanting physicians.

Root Cause

Packaging process control

Distribution Pattern

Worldwide: USA and Australia