Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

UltraLite Vehicles powered by Rascal Motorized Vehicles.

FDA Recall Z-1376-05 · Posted 2005-08-17

Recall StatusTerminated
Product CodeITI
Date Initiated2005-05-05
Date Terminated2006-10-03
Product Quantity788 units

Reason for Recall

The engager set screw may become loose. This problem may cause the potentiometer to become off-center thus affecting the end users stopping capabilities.

Firm Action

The firm sent out letters to their customers on 4/20/2005. The firm's sales representatives were also notified on 4/20/2005.

Root Cause

Other

Distribution Pattern

The firm sells the vehicles via direct sales to end customers and also to company owned inventory with sales representatives. The distribution is nationwide.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑