Radius Rod Persuader is a non-sterile instrument for use with other devices in spinal surgery.

FDA Recall Z-1377-2008 · Posted 2008-08-19

Recalling FirmStryker Spine
Recall StatusTerminated
Product CodeLXH
Date Initiated2008-02-19
Date Terminated2008-08-19
Product Quantity65 units

Reason for Recall

The t-bar of the Radius Rod Persuader devices froze due to a loosening screw causing the device to fail.

Firm Action

Recall notification was sent via FedEx on February 19, 2008 to all branches/agencies. All questions are to be directed to Michelle Klein, Regulatory Compliance Specialist - 201-760-8256.

Root Cause

Device Design

Distribution Pattern

Product was distributed to a limited population during a product launch for the purposes of gathering feedback on the product.