Radius Capspin is a non-sterile instrument for use with other devices in spinal surgery.
FDA Recall Z-1378-2008 · Posted 2008-08-19
Advertisement
Recalling FirmStryker Spine
Recall StatusTerminated
Product CodeLXH
Date Initiated2008-02-19
Date Terminated2008-08-19
Product Quantity52
Reason for Recall
The Radius Capsin has separation of the tip from the inserter.
Firm Action
Recall notification was sent via FedEx on February 19, 2008 to all branches/agencies. All questions are to be directed to Michelle Klein, Regulatory Compliance Specialist - 201-760-8256.
Root Cause
Device Design
Distribution Pattern
Product was distributed to a limited population during a product launch for the purposes of gathering feedback on the product.