CS 8100 3D, CATALOG # (s): 5311162, 5311188. Intended to produce complete or segmented tomographic digital panoramic X-ray images and three-dimensional digital X-ray images of the dentomaxillo-facial area to be used at the direction of healthcare professionals as diagnostic support for pediatric and adult patient.

FDA Recall Z-1378-2015 · Posted 2015-04-03

Recall StatusTerminated
Product CodeMUH
Date Initiated2015-01-08
Date Terminated2017-04-20
Product Quantity470 units (127 domestically & 343 internationally)

Reason for Recall

Units device head descended unexpectedly

Firm Action

The firm sent customers "Urgent: Medical Device Recall" letters, dated December 29, 2014. The letter described the problem, as well as, the actions to be taken. A Carestream Health service representative will contact customers to schedule a visit to inspect the affected devices. Questions or concerns should be directed towards the Carestream Customer Care Center in the U.S. at 1-800-328-2910.

Root Cause

Component design/selection

Distribution Pattern

Worldwide Distribution -- US, including the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NV, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VT, WA, WI, and WV; and the countries of India and Switzerland.

Official Source

View FDA Recall →