Spetzler Claw Tip, Universal - 5450-800-311 Spetzler Open Angle Micro Claw Tip, Universal - 5450-800-313 The Spetzler Claw" Tips are bone cutting tips which are a part of Strykers Neuro Spine Ear, Nose and Throat (ENT) Sonopet Portfolio. They are removable attachments intended to be used with the Sonopet Ultrasonic Handpiece.

FDA Recall Z-1378-2016 · Posted 2016-04-13

Recall StatusTerminated
Product CodeLFL
Date Initiated2016-03-22
Date Terminated2016-07-19
Product Quantity140 units

Reason for Recall

The Spetzler Claw Tip was packaged with the incorrect Tip Sleeve Assembly. The Tip Sleeve Assembly included in the packaging is the incorrect size. Risk to Health: If the Tip Sleeve Assembly does not fit the Spetzler Claw" Tip, a surgical delay may occur while an alternate Tip Sleeve Assembly is retrieved.

Firm Action

On 3/22/2016, URGENT MEDICAL DEVICE RECALL NOTIFICATION notifications were sent to the affected customers via courier. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. For questions regarding this recall, please contact Stryker Instruments: Kara Spath 269-389-4518 [email protected]

Root Cause

Mixed-up of materials/components

Distribution Pattern

Distributed in the states of AZ, CA, CO, CT, IL, IN, MD, MI, MN, NY, PA, SC, TN, WA, WI, and VA, and in the NETHERLANDS.