Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

Modular Knee Joint with Friction Brake

FDA Recall Z-1379-04 · Posted 2004-08-18

Recall StatusTerminated
Product CodeISH
Date Initiated2004-07-15
Date Terminated2005-11-07
Product Quantity457 devices

Reason for Recall

Braking Plates of the 3R90 and 3R92 Modular Knee Joint with Friction Brake may become dislodged which may cause the braking performance of the knee to malfunction.

Firm Action

The braking plate of the knee may become dislodged, which may cause the braking performance of the knee to malfunction.

Root Cause

Other

Official Source

View FDA Recall →