Terminated

Accu-Chek Advantage Plus blood glucose test strips; International Reference no. 04735102001.

FDA Recall Z-1382-06 · Posted 2006-08-11

Recall StatusTerminated
Product CodeLFR
Date Initiated2006-06-27
Date Terminated2007-05-04
Product Quantity21,000,000 consumer vials and 10,000,000 hospital vials to U.S. plus 35,000,000 vials international-(Totals for Recalls Z-1377-1384-06)

Reason for Recall

The drying agent beads, located in the cap, may become loose and fall into the vial causing the test strips to potentially give incorrect blood glucose test results

Firm Action

U.S. consignees, point of care coordinators or laboratory managers, long term care professionals, health care professionals, pharmacists, and mail order distributors were notified via letter dated 6/28/06 to cease using the strips and to contact the firm for replacement if they see loose beads in their vial. International accounts were also notified.

Root Cause

Other

Distribution Pattern

Worldwide Distribution-USA and countries of Argentina, Australia, Brazil, Canada, Chile, China, Colombia, Ecuador, England, France, Germany, Hong Kong, India, Indonesia, Japan, Malaysia, Mexico, New Zealand, Pakistan, Paraguay, Peru, Philippines, Singapore, South Africa, South Korea, Spain, Sweden, Taiwan, Thailand and Uruguay.