Terminated

Siemens Axiom Luminos dRF solid state x-ray imager Solid state x-ray imager (flat panel/dig. imager)

FDA Recall Z-1383-2012 · Posted 2012-04-09

Recall StatusTerminated
Product CodeMQB
Date Initiated2012-02-16
Date Terminated2014-11-06
Product Quantity2

Reason for Recall

During regular product monitoring, Siemens became aware of a potential issue with the AXIOM Luminos dRF. Firm became aware of a potential issue with torque wrenches - during a regular scheduled calibration of the wrench, it was discovered that the wrench was out of the specification. The wrench was over-torquing approximately 21% for low-end readings (50 ftlbs) and 9% for high-end readings (250

Firm Action

Siemens initiated a recall/correction on February 16, 2012 by issuing Update Instruction XP056/11/S to the affected customers. All screws on the unit support on AXIOM Luminos dRF units will be replaced by a Siemens service engineer. For information regarding this recall call 610-219-4834.

Root Cause

Nonconforming Material/Component

Distribution Pattern

Nationwide Distribution including California

Related FDA 510(k) Records