Terminated

The Model 8540 Catheter Access Port Kit is intended for use in accessing the catheter via the catheter access port of Medtronic(R) implantable programmable infusion pumps (except Medtronic MiniMed pumps). The Model 8540 Catheter Access Port Kit does not have a product code for itself; it is approved under PMA P860004. PMA P860004 is classified as: pump, infusion, implanted, programmable under product code LKK.

FDA Recall Z-1384-04 · Posted 2004-08-24

Recall StatusTerminated
Product CodeLKK
Date Initiated2004-07-23
Date Terminated2005-11-23
Product Quantity405 kits

Reason for Recall

The sterile tray label for the Model 8540 Catheter Access Port kit may be incorrectly labeled with a Model 8551 Refill kit label. The outer box of the affected Model 8540 Catheter Access Port kit is labeled correctly and the component parts within the sterile tray are the correct components for a Model 8540 Catheter Access Port kit. The issue is limited to one lot of product: CS0293.

Firm Action

The sterile tray label for the Model 8540 Catheter Access Port kit may be incorrectly labeled with a Model 8551 Refill kit label. The outer box of the affected Model 8540 catheter Access Port kit is labeled correctly and the component parts within the sterile tray are the correct components for a Model 8540 Catheter Access Port kit. The issue is limited to one lot of manufactured product: CS0293.

Root Cause

Other

Distribution Pattern

US and one customer in New Zealand.