EMPOWR Porous Knee Patella Over Drill, REF: 802-05-104; EMPOWR Knee Patella Drill, REF: 802-05-108
FDA Recall Z-1384-2023 · Posted 2023-04-12
Advertisement
Recalling FirmEncore Medical, LP
Recall StatusOpen, Classified
Product CodeMBH
Date Initiated2023-02-28
Product Quantity29
Reason for Recall
Patella drills are marked with incorrect diameter size of 5.6mm, when they should be 5.4mm. Porous patella overdrills are marked with incorrect diameter size of 5.4 mm, when they should be 5.6 mm. If the incorrect drill is selected during surgery, there is risk of the implant loosening, implant not properly fitting, and/or fractured bone.
Firm Action
On 2/28/23, field safety notices were emailed to surgical sales agents who were asked to document the quantity of affected devices complete and return the acknowledgement form. Customer service contact, [email protected], was provided to report failures or to request replacements. Customers with questions can phone (914) 450-3931, or email [email protected].
Root Cause
Process change control
Distribution Pattern
US: TX, IN, MD, CA, KS, AL, HI, WA
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.