CARDIOSAVE Hybrid IABP and CARDIOSAVE Rescue IABP affected part number is 0998-:XX-0800-XX Li-Ion Battery affected part number is 0146-00-0097 Product Usage: The CARDIOSAVE Hybrid Intra-Aortic Balloon Pump (IABP) is an electromechanical system used to inflate and deflate intra-aortic balloons

FDA Recall Z-1386-2017 · Posted None

Recall StatusTerminated
Product CodeDSP
Date Initiated2016-12-23
Date Terminated2019-03-01
Product Quantity8000

Reason for Recall

Maquet has received 3 confirmed complaints related to the Li-lon Battery used with the Maquet CARDIOSAVE Hybrid IABP and Maquet CARDIOSAVE Rescue IABP of the battery venting after being accidentally dropped.

Firm Action

Maquet contacted their customers via letter on 12/23/16 and additional customers identified were notified on 1/10/17 and issued a recall letter/return response form. The letter identified the affected product problem and actions to be taken. Customers were instructed to complete and return the response form via fax to 1-973-629-1518 or email to [email protected]

Root Cause

Device Design

Distribution Pattern

Worldwide Distribution - US Nationwide