SOUTHERN IMPLANTS REF IBR12d-13, Implant, Ext. Hex, tapered 12o Co-Axis(TM) intended for surgical placement in the upper or lower jaw to provide a means for prosthetic attachment of crowns, bridges or overdentures utilizing delayed or immediate loading.

FDA Recall Z-1386-2018 · Posted None

Recall StatusTerminated
Product CodeNHA
Date Initiated2018-03-08
Date Terminated2019-09-26
Product Quantity49 units

Reason for Recall

The UDI label correctly reads IBR12d-13, however, the inner label incorrectly read IBNT12d-13.

Firm Action

The firm initiated their recall by email on 03/08/2018. The firm requested return of the product.

Root Cause

Process control

Distribution Pattern

US, South Africa, France