Blood Gas and Co-Ox Electrolyte & Metabolyte Analyzers, ABL-800 series, manufactured from 8/2004 through 9/2005.

FDA Recall Z-1387-06 · Posted 2006-08-10

Recall StatusTerminated
Product CodeJFP
Date Initiated2005-09-29
Date Terminated2012-04-17
Product Quantity196 devices

Reason for Recall

Under certain circumstances, the ABL 700 &800 Series Blood Gas Analyzers may experience leakage current into the measuring system. Consequently, the analyzers intermittently provide incorrect (too low) result values for Calcium (Ca) and Sodium (Na).

Firm Action

The consignees were notified of this recall by a letter, dated 9/29/2005. The letter informs the customers of this potential problem and the availability of new updated software which will correct the problem. The letter goes on to inform the customer that a Radiometer Service Engineer wil be contacting them shortly in order to arrange for a convenient time to install the software update. Along with the letter, the customer is provided a copy of an additional page and a CD concerning the function and operation of the new software, which is to be added to the customer''s Operator''s Manual.

Root Cause

Other

Distribution Pattern

Nationwide.