Access Immunoassay Systems Assay Protocol: Dil-AFP.
FDA Recall Z-1388-05 · Posted 2005-08-23
Advertisement
Recalling FirmBeckman Coulter Inc
Recall StatusTerminated
Product CodeLOJ
Date Initiated2005-07-18
Date Terminated2012-03-07
Product Quantity311 USA 16 Canada
Reason for Recall
Pipetting for dil-Alpha-fetoprotein when performed on the Access, Access 2 and Synchron LX i 725 has a potential to create splashing which may cause erroneous results.
Firm Action
Letters to customers were sent July 18, 2005. Firm will upgrade software later in the year. Customers are advised to not perform the automated dilution protocol but refer to the method in the directions.
Root Cause
Other
Distribution Pattern
Nationwide and Canada
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.