Human Fetal Hemoglobin (HBF-1) FITC Conjugate

FDA Recall Z-1388-2018 · Posted None

Recall StatusTerminated
Product CodeGHQ
Date Initiated2018-02-07
Date Terminated2019-05-08

Reason for Recall

Three lots of Human Fetal Hemoglobin FITC Conjugate are displaying high background and unusual bright staining in the controls and patient samples which could potentially lead to erroneous results (i.e. false positives)

Firm Action

All consignees which have purchased the Human Fetal Hemoglobin FITC antibody including facilities/customers that have shipped the lots outside of their facility will be notified via certified mail to check inventory to see if affected product is on hand. Consignees instructed to discontinue use and discard per internal procedure. The effectiveness check will begin 10 days after the distribution date of the Medical Device Recall Notification. The field removal is considered effective when 10% of the affected customers acknowledge receipt and understanding of the notification instructions.

Root Cause

Use error

Distribution Pattern

MI, CA, IL, TX, SC, NY, MN, DE, NJ, WA , OR, IN, MA, DE, FL, PA

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.