The GMD UNIVERSAL URINARY INCONTINENCE SLING (Code 1010) is supplied sterile and is intended for single patient use only. The GMD Universal Sling is a sterile, single use device for the treatment of female stress urinary incontinence. It is comprised of a polypropylene sheath with a disposable universal low density polyethylene sleeve at each end of attachment of the sling to GMD's single use or reusable trocars. The universal sleeve has three trocar insertion points, the distal and proximal trocar insertion points are for inside-out, bottom-up approaches and a sleeve end trocar insertion point is for outside-in, top-down approaches. The design allows the physician to utilize the device for placement of a polypropylene mesh at the urethra or bladder neck.

FDA Recall Z-1390-2011 · Posted 2011-02-22

Recall StatusTerminated
Product CodeFTL
Date Initiated2011-02-14
Date Terminated2012-09-24
Product Quantityapproximately 1711

Reason for Recall

Protective sheath sticks to itself and sleeve breakage has resulted in complaints and MDRs.

Firm Action

The firm, GMD, issued an "URGENT MEDICAL DEVICE RECALL" letter dated February 14, 2011 to all customers. The letter described the product, problem and action to be taken by the customers. The customers were instructed not to use any GMD Model 1010 Universal Slings remaining in their inventory; return any Model 1010 slings in their possession to the Distributors or GMD; notify their customers to whom they shipped Model 1010s of this voluntary recall and request them to return any Model 1010 slings in their possession to them; and return all Model 1010 inventories to GMD using the attached Recalled Model 1010 Return Form. Note: Full credit in the amount of the returned products purchase price will be issued for all sterile, unexpired product returned to GMD. Contact the GMD Product Performance Senior Manager, at phone: (253) 853 3512 or email: [email protected], if you have any questions/issues related to this recall.

Root Cause

Component design/selection

Distribution Pattern

Worldwide distribution: USA and countries including: Australia, Argentina, Brazil, Canada, South Africa, and Scandinavia.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.