EasyLink(TM) Informatics Systems Software Version 5.0 and Software Version 5.0 Service Packs 1 through 4 used with the Dimension Vista(R) System. The EasyLink(TM) Informatics System is classified as a data calculator and processing module for use with In-vitro diagnostics devices. The EasyLink(TM) is included with the Dimension Vista(R) System, and may be used with other Siemens analyzers, as a communications and connectivity workstation for integration with laboratory information system (LIS) networks. Product Quantity Distributed (Int) 787

FDA Recall Z-1391-2012 · Posted 2012-04-10

Recall StatusTerminated
Product CodeJJE
Date Initiated2011-12-02
Date Terminated2014-03-21
Product Quantity787

Reason for Recall

QC results may be unexpectedly associated with a Pending Control. When this situation occurs, Pending QC Controls maybe be created or reused.

Firm Action

Siemens Healthcare Diagnostic Inc, sent a Urgent Field Safety Notice dated December 2011, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers were asked to follow these instructions. The first option is Extending the "Minimum Inactive Days" to 999 days. The second option is to proactively monitor the system for unexpected Pending QC Controls. Long Term Action Plan: A corrective and preventive action (CAPA) investigation has been issued. Currently firm is actively investigating. Further questions please call (302) 631-0516.

Root Cause

Device Design

Distribution Pattern

Worldwide Distribution -- USA (nationwide) and the countries of CANADA, FRANCE, GERMANY, SPAIN, PORTUGAL, ITALY, JAPAN, NORWAY, AUSTRALIA, AUSTRIA, NEW ZEALAND, BELGIUM, SWITZERLAND, NETHERLANDS, MALAYASIA, SLOVENIA, SOUTH KOREA, DENMARK, SAUDI ARABIA, SLOVAKIA, PUERTO RICO, INDONESIA and TAIWAN.