SMV FX-40 Nuclear Camera
FDA Recall Z-1393-04 · Posted 2004-08-24
Advertisement
Recalling FirmGeneral Electric Med Systems
Recall StatusTerminated
Product CodeIYX
Date Initiated2004-07-30
Date Terminated2006-07-24
Product Quantity65
Reason for Recall
The detector head of a FX-80 Nuclear camera may fall from the gantry. Three potential causes have been determined. (1) A ball nut in the detector's radial drive assembly can back out of the bearing block that contains it. (2) Misalignment of the radial drive assembly. (3) Mounting bolts on the radial drive assembly may become loose, fail, or back out.
Firm Action
The detector head of a FX-80 Nuclear camera may fall from the gantry. Three potential causes have been determined. (1) A ball nut in the detector''s radial drive assembly can back out of the bearing block that contains it. (2) Misalignment of the radial drive assembly. (3) Mounting bolts on the radial drive assembly may become lose, fail, or back out.
Root Cause
Other
Distribution Pattern
Inside the US. Manufactured and distributed from 1993 - 2000.