RTI Surgical Streamline Occipito-Cervico-Thoracic (OCT) Final Driver, Common Name: Screw Driver
FDA Recall Z-1393-2016 · Posted None
Reason for Recall
Multiple lots of the 26-FNLDRIVER, may have reduced mechanical strength, resulting in driver tip fracturing during use.
Firm Action
RTI Surgical Inc. initiated a voluntary recall of RTI Surgical Streamline OCT Occipito-Cervico-Thoracic Final Driver, via written notices sent by e-mail and FedEx on 03/18/2016.
Customers are requested to review all inventory, quarantine products listed in the recall, and to immediately return recalled products with the recall reply form, and the shipping labels and authorization numbers provided within 5 days of receiving notice. Replacement drivers are being sent to customers.
Customers may contact Dan Nelson, Manager of Quality Systems, Phone: 906-226-4489, Fax: 906-226-4455, e-mail: [email protected].
Root Cause
Process design
Distribution Pattern
Domestic: AL, CO, CT, FL, IN, ID, IL, MI, MS, MO, NC, NY, TX Foreign: None VA/DOD: None
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.