Maquet Cardiopulmonary AG, MAQUET RPM 20: Roller Pump for Heart Lung Machine HL20.
FDA Recall Z-1394-2011 · Posted 2011-02-22
Advertisement
Recalling FirmMaquet Cardiovascular Us Sales, Llc
Recall StatusTerminated
Product CodeDTQ
Date Initiated2010-09-28
Date Terminated2011-08-11
Product Quantity4 units
Reason for Recall
A wrong ball bearing was installed in the pump heads of product. This may make the pump head wear out and cease function.
Firm Action
Maquet hand delivered a letter to the consignee on 10/4/2010.
Root Cause
Component change control
Distribution Pattern
One hospital in New York City