Terminated

BD AffirM VPIII Microbial Identification Test

FDA Recall Z-1394-2017 · Posted 2017-02-27

Recall StatusTerminated
Product CodeMLA
Date Initiated2015-08-27
Date Terminated2017-03-06
Product QuantityQty Distributed to Field - 4,502

Reason for Recall

BD has confirmed that a portion of tests associated with the affected lots( 446252 and 446257 ) of product may exhibit blue speckles or fibers on the beads and cards, as shown below. This issue would result in an invalid test, as described in the package insert.

Firm Action

Becton Dickinson sent an Urgent Product Recall Letter dated August 2015. The distributors were provided with a written communication via email and customers were provided with a written communication via UPS. Customers were not asked to send product back. If further assistance is needed regarding replacements, please contact BD Customer Service Department at 1-800-675-0908. For all other inquiries, please contact BD Technical Services Department at 1-800-638-8663. For further questions please call (410) 316-4428.

Root Cause

Employee error

Distribution Pattern

Worldwide Distribution - USA (nationwide) Distribution including Puerto Rico and to the states of : AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV and WY and to the countries of : AU, BE, IN, KR, SG,

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.