Terminated

Brilliance 64 with 4.1.7 XX026 software version model number 728231 Product Usage: Computed Tomography X-ray systems intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes. These devices may include signal analysis and display equipment, patient and equipment supports, components, and accessories

FDA Recall Z-1396-2018 · Posted None

Recall StatusTerminated
Product CodeJAK
Date Initiated2018-02-21
Date Terminated2020-04-10
Product Quantity291

Reason for Recall

Surview scan with tube current lower than 30 mA is unable to be initialized.

Firm Action

On February 21, 2018 Philips Medical Systems sent a Customer Information Letter regarding the Philips Brilliance 64 Software Model 4.1.7. Customers were informed of how to identify affected products and instructed to use the Instructions For Use recommended 30mA for surview scans. Customers with questions or concerns should contact their local Philips representative or local Philips Healthcare office Customer Care Solutions Center at 1-800- 722-9377.

Root Cause

Under Investigation by firm

Distribution Pattern

Worldwide Distribution - US Nationwide in the states of GA, VA, MI, AZ, and the countries of Canada, Switzerland

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.