CVAC Aspiration System, REF: CVC127020-1, and User Manual L00018. Intended for endoscopic examination/treatment of urinary tract and kidney interior.

FDA Recall Z-1396-2025 · Posted 2025-03-21

Recalling FirmCalyxo, Inc.
Recall StatusOpen, Classified
Product CodeFED
Date Initiated2025-02-19
Product Quantity11,246

Reason for Recall

Aspiration system, for endoscopic examination/treatment of urinary tract and kidney interior, to have labeling update adding additional instructions, for patients with high viscosity fluid in the kidney, not to continue providing fluid inflow in the presence of unresolved slow or absent fluid outflow because this can create intrarenal pressure imbalance, and lead to excessive intrarenal pressure.

Firm Action

On 2/19/2025, correction notices were mailed to distributors and healthcare providers who were asked to do the following: 1) Notify all urologists utilizing the affected aspiration system in your institution that additional instructions are required. Please provide each of them with a copy of the additional instructions. 2) As another safety step, a firm representative will be contacting all users from your institution directly to offer training on these additional instructions. If you prefer to coordinate the training, please specifically state that preference in the acknowledgement. 3) Please acknowledge receipt of the notice, additional instructions, and updated training instructions MA00100.A by emailing [email protected], or calling 833-214-3354 and providing your name, title, institution and address.

Root Cause

Device Design

Distribution Pattern

US Nationwide distribution including in the states of AL, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, OH, OK, OR, PA, SC, TN, TX, UT, VA, VT, WA, WI, WV.