ONCOR Avant-garde, Digital Linear Accelerator, Model No. 05863472 Product Usage: The intended use of the device is to deliver x-ray radiation for therapeutic treatment of cancer.

FDA Recall Z-1398-2018 · Posted None

Recall StatusTerminated
Product CodeIYE
Date Initiated2018-02-14
Date Terminated2019-05-16
Product Quantity1

Reason for Recall

Control Console software has been updated to reduce the risk for collision when delivering automatically sequenced treatments with automatic movements of the gantry and/or the treatment table.

Firm Action

On February 16, 2018, the recalling firm issued a Customer Advisory Notice via USPS Certified Mail to inform customers about the update of the Control Console software version 9.2.400 of the Digital Linear Accelerator to version 9.2.502. Customers were informed of the issues that would be fixed through this software update. Consignees were asked to instruct affected personnel accordingly, place the safety advisory notice in the device's instructions for use, and have personnel maintain awareness until the fix is implemented.

Root Cause

Software design

Distribution Pattern

Worldwide Distribution- US Nationwide in the states: AL, CA, FL, GA, IL, IN, KY, MA, MI, MO, MS, NJ, NY, OH, OK, PA, RI, SC, TN, TX, WI, WV, and WY; and PR. International distribution to Nassau.