GammaMed software program, version 5.07, vers. 5.08, vers.-usa, vers.-tst and vers.-row , part number GM11005400, for the GammaMed model 12it radionuclide applicator system.
FDA Recall Z-1399-06 · Posted 2006-08-10
Recalling FirmVarian Medical Systems
Recall StatusTerminated
Product CodeJAQ
Date Initiated2006-05-01
Date Terminated2006-12-05
Product Quantity13 units
Reason for Recall
Software control program for a medical device used in radiation treatment may cause practitioners to incorrectly administer the treatment plan to cancer patients. There is a hazard when entering the treatment plans manually that the user neglects to change the default step size and/or the origin or accidentally enters incorrect parameters.
Firm Action
The recalling firm notified healthcare practitioners using the medical device by letter on 4/17/06 and continuing through 5/1/06. The notification was flagged as a 'Medical Device Recall' and instructed users to adjust system parameters in the autoexec.bat file. The notification further recommended using the identified treatment plans and that manual input should be used for QA purposes only.
Root Cause
Other
Distribution Pattern
Nationwide.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.