Terminated

Immulite 2000 and Immulite 2500 digital Fluidics PC Board, Part Number 450116-0002. Performs chemiluminescent immunoassays using serum, plasma, and/or urine for in vitro diagnostic testing.

FDA Recall Z-1399-2009 · Posted 2009-05-26

Recall StatusTerminated
Product CodeDGC
Date Initiated2009-01-05
Date Terminated2010-04-19
Product Quantity295 Digital Fludics PCB (88 IMMULITE 2000 and 2500 and 207 as spare parts).

Reason for Recall

Design Flaw: The redesign of the Digital Fluidics PCB used in IMMULITE 2000 and IMMULITE 2500 is missing a jumper.

Firm Action

Siemens Healthcare Diagnostics Inc. issued an "Urgent: Device Recall" notice dated January 2009 mailed to all direct customers of the affected product. The letter described the problem, provided "work-around" instructions and also provided a Recall Response Form to be returned to the firm via fax. For further information, please contact your local Siemens Healthcare support representative or Siemens Healthcare Diagnostics Inc. directly at 1-973-927-2828.

Root Cause

Process design

Distribution Pattern

Worldwide Distribution to states including MA, NY, PA, MD, GA, FL, OH, IN, MI, IL, MO, LA, UT, AZ and CA and countries of Canada, Czech Republic, Germany, Poland, Hungary, Belgium, Russia, Republic of South Africa, Greece, Israel, Turkey, Spain, Portugal, Denmark, France, Great Britain, India, China, South Korea and Mexico.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.