BD Phoenix Yeast ID Panel is a self-inoculating (when used with a BD Phoenix Automated Microbiology System) polystyrene panel (tray) containing control wells and substrate wells with dried biochemical substrates that use traditional qualitative microbiology methods such as fermentation, oxidation, degradation, and hydrolysis in combination with chromogenic and fluorogenic substrates along with carbon sources for yeast identification.
FDA Recall Z-1399-2024 · Posted 2024-03-29
Advertisement
Recalling FirmBecton Dickinson & Co.
Recall StatusOpen, Classified
Product CodeLON
Date Initiated2024-02-29
Product Quantity6700 eaches
Reason for Recall
Esculin, a substrate used in other panels and
normally in A16 well, was found in A17 well. A17 in yeast panels is supposed to be an
empty well and is not a part of the normal ID evaluation. However, A17 is checked for
fluorescence and if anything is found, the system will abort the panel with the message
Fluorescent interference has been detected. Esculin is expected to trigger this
fluorescence and test cancellation.
Firm Action
BD issued URGENT: Medical Device Product Correction Product Removal letter on 2/29/24 via courier service via tracked letter and/or e-mail. Letter states reason for recall, health risk and action to take:
1. Immediately inspect your inventory for the specific catalog and lot number listed above. Destroy affected product subject to the recall following your institution s process of destruction.
2. Ensure the contents of this notification are read and understood.
3. Share and post this customer letter within your facility network and forward to any customers to whom you may have distributed the product to ensure awareness.
4. Complete the attached Customer Response Form and return to the BD contact noted on the form whether or not you have any of the impacted material so that BD may acknowledge your receipt of this notification per FDA requirements.
BD will issue product replacements following receipt of the completed Customer Response Form.
Contact: North American Regional Complaint Center
Phone: 1-844-8BD-LIFE (1-844-823-5433)
Say Recall when prompted Mon Fri 8:00am and 5:00pm CT
or
Email: [email protected]
Product Advisory, Recall Questions, Product Complaints, Technical Questions
Root Cause
Under Investigation by firm
Distribution Pattern
Worldwide distribution - US Nationwide and the countries of AR, BE, CO, IN, JP, PE, PY, SG, TW, UY.