Terminated

Remel, Haemophilus Test Medium (Agar) in A) 100 mm REF R01503, and B) 150 mm plates REF R04033,

FDA Recall Z-1400-2017 · Posted None

Recalling FirmRemel Inc
Recall StatusTerminated
Product CodeLKA
Date Initiated2017-02-09
Date Terminated2019-05-01
Product QuantityRO4033 - 163, R01503 - 50.

Reason for Recall

Test organisms exhibit poor growth when grown using the test agar. The poor growth can give users incorrect AST zones when performing susceptibility tests.

Firm Action

Thermo Fisher sent an Urgent Medical Device Recall letter dated February 9, 2017, to all affected customers. The recalling firm issued written notifications to their customers. Customers were instructed to destroy any remaining inventory of the recalled lots. Customers were also instructed to contact everyone within their organization or any other organization where the devices have been transferred. An acknowledgement form was included with the recall notice. Customers were asked to complete the acknowledgement form and return it to the recalling firm. If you have any questions, please contact our Technical Service Department at (800) 255-6730 ( US ) or (913) 888-0939 ( International) For further questions, please call (913) 895-4077.

Root Cause

Nonconforming Material/Component

Distribution Pattern