Canon Full Auto Ref-Keratometer RK-F2, Catalog Number 6937B002AA. Product Usage: This device is intended to be used to measure the refractive power and the radius of corneal curvature of the human eye.
FDA Recall Z-1404-2013 · Posted 2013-05-25
Advertisement
Recalling FirmCanon Inc.
Recall StatusTerminated
Product CodeHKO
Date Initiated2012-12-05
Date Terminated2013-09-20
Product Quantity384 units
Reason for Recall
Canon Inc. had received 17 claims from our foreign
sales companies between October to November
2012 that measurement value of RK-F2s refractive
power (SPH value) sometimes indicated abnormal
value which occurs approximately 5% in probability.
Firm Action
Canon initiated recall by sending letters on November 22, 2012 to all affected customers. The letter identified the affected product, problem, actions to be taken and contact information.
Root Cause
Software design
Distribution Pattern
Worldwide Distribution - USA Nationwide and countries of: Australia, Japan, Korea, Singapore and United Kingdom.