Terminated

AngioDynamics Micro Introducer Kits under the following labeling: 1) AngioDynamics 4F Standard Micro-Introducer Kit, Sterile, Catalog Number: 06597005, 2) AngioDynamics 5F Standard Micro-Introducer Kit, Sterile, Catalog number: 06597012; 3) AngioDynamics 4F Standard Micro-Introducer Kit, Sterile, Catalog Number: 06597013; 4) AngioDynamics 5F Standard Micro- Introducer Kit, Sterile, Catalog Number: 06597018; and 5) AngioDynamics 4F Standard Micro-Introducer Kit, Sterile, Catalog Number: 06597019. Product kits are individually wrapped in plastic, 10 units included per box. Product Usage: This device is intended to be used to measure the refractive power and the radius of corneal curvature of the human eye.

FDA Recall Z-1405-2013 · Posted 2013-05-25

Recalling FirmAngiodynamics, Inc.
Recall StatusTerminated
Product CodeDQR
Date Initiated2013-04-30
Date Terminated2016-11-09
Product QuantityDomestic(1,760 units): 1)260 units/26 boxes, 2)310 units/31 boxes,3)260 units/26 boxes,4)880 units/88 boxes,5)60 units/6 boxes,Foreign:10 units/1 box

Reason for Recall

AngioDynamics Inc., is recalling Micro Introducer Kits and VenaCure EVLT NeverTouch Procedure Kits, because the potential exists that the micro Introducer (sheath/dilator) packaged within the kit could be the incorrect French size.

Firm Action

AngioDynamics sent an Urgent Medical Device Recall letter dated April 29, 2013, to all customers by Fed Ex. The letter identified the affected product, problem and actions to be taken. Customers were instructed to lmmediately remove the affected product from inventory and call AngioDynamics, Inc. Customer Service at 1-800-772-6446 to obtain a replacement or credit for your returned product and to obtain UPS Account Number to return the package via second day delivery. Customers are instructed to segregate and return all affected devices to: AngioDynamics 603 Queensbury Avenue Queensbury, NY 12804 Attn: Micro Introducer and VenaCure EVLT NeverTouch Procedure Kits Recall Coordinator. Write the RMA number on the shipping box (Provied on the Recall Verification Tracking Form). Promptly complete, sign and return the enclosed Reply Verification Tracking Form (even if you do not have any product to return); following the directions on the notification form and the Reply Verification Form. Send the Reply Verification form to AngioDynamics either via Fax or Emai. Fax Reply Verification Tracking Form to: Attn: Micro Introducer and VenaCure EVLT NeverTouch Procedure Kits Recall Coordinatory Fax number 1-518-798-1360 or Email Reply Verification Tracking Form to: [email protected].

Root Cause

Labeling mix-ups

Distribution Pattern

USA Nationwide Distribution including the states of: IL, GA, CA, NY OH, AZ, TX, NE, FL, PA, NC, WI, AL, IN, CO, MO, MT, NE, MA, VA, AK, IN, MI, NJ, MD, TN, CT, WA, MN