Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

A-series lift - A625 Manual w/Load Cell & Wi-Fi, Omni Part No. 324124. Prism Medical.

FDA Recall Z-1405-2017 · Posted 2017-02-28

Recall StatusTerminated
Product CodeFSA
Date Initiated2017-01-30
Date Terminated2017-12-04
Product Quantity21

Reason for Recall

During internal cycle test at max load (625 lbs.) 2 units had straps break prematurely.

Firm Action

The recalling firm called each customer on Jan 30, 2017 to review the recall and schedule a time for the firm to come to their facility to personally swap the recalled product for a new device. The FSNs were sent on Jan 31, 2017.

Root Cause

Device Design

Distribution Pattern

Product was distributed to MT and NV; one VA consignee.

Official Source

View FDA Recall →