Prodisc C - prosthesis, intervertebral disc - Product Usage: indicated in skeletally mature patients for reconstruction of the disc from C3-C7 following single-level discectomy for intractable symptomatic cervical disc disease (SCDD), Part number: 09.820.055S
FDA Recall Z-1406-2021 · Posted None
Advertisement
Recalling FirmCentinel Spine, Inc.
Recall StatusTerminated
Product CodeMJO
Date Initiated2021-03-05
Date Terminated2021-08-02
Product Quantity49 devices
Reason for Recall
Centinel Spine learned through five customer complaints that the product was missing the "UP" etching on the superior plate of the prodisc C device. Per the Surgical Technique, the "UP" is one of the methods used to indicate the direction in which the surgeon should place the prodisc C into intradiscal space. There are also "UP" indications in the packaging and on instrumentation as additional guides to help the surgeon properly implant the device that were present for the affected product.
Firm Action
In March 2021, Centinel issued an Urgent Medical Device Recall notice to customers and distributers via letter and email notifying them the "UP" etching on the superior plate is missing. The "UP" is used to signify that the implant should be inserted into the intradiscal space with "UP" towards the head or in the cephalic position. In those devices with "UP" missing, there is a risk that the device could be placed upside-down within the intradiscal space.
Root Cause
Employee error
Distribution Pattern
US Nationwide distribution in the states of CA, MD, DC, TN, GA, NC, PA, KY, NY, AZ, CO, OH, FL, MI.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.