INOvent, Nitric Oxide Delivery System. Model Number 1605-9000-000

FDA Recall Z-1410-06 · Posted 2006-08-25

Recall StatusTerminated
Product CodeMRN
Date Initiated2006-06-29
Date Terminated2007-07-16
Product Quantity3,123 units

Reason for Recall

The Kel-F Tip mounted on the high pressure hose can become dislodge/lodged in the INOmax valve outlet.

Firm Action

On June 29, 2006, the firm sent a letter to its consignees advising them to visually verify that the Kel F tip is in place before attempting to connect either of the high pressure hoses to the INOmax cylinders. INO advised that if the tip was not in place, the operator should immediately replace the tip before attaching the hoses. On July 5, 2005, the firm sent a second letter to their consignees with extra Kel F tips and existing instructions from the INOvent Operation & Maintenance Manual outlining the proper replacement procedure for the tips. The firm is following with telephone contact to assure that the letters were delivered to the proper person(s).

Root Cause

Other

Distribution Pattern

Worldwide, USA and countries of Canada, Uruguay, Chile, Argentina, Colombia, and Malaysia